The x-ray measurements described in this, and subsequent
chapters were all performed upon the same triple-crystal
diffractometer, using a GEC Avionics GX21 rotating anode x-ray
source. The rotating anode operates at 3kW on a copper target
producing CuK
radiation. A schematic picture of the
triple-crystal arrangement is shown in Figure 3.6,
illustrating the scattering geometry of monochromator,
sample, and analyser crystals. The CuK
wavelength,
is selected by the appropriate
adjustment of monochromator and source-slit. A further
adjustable pre-sample slit is used to reduce both vertical
and horizontal beam size to the appropriate dimensions of
the sample. The detector is a proportional counter.
The use of an analyser crystal ensures a resolution which is
limited only by the mosaic spread of the set of crystals chosen
for monochromator and analyser. This has the advantage of
discriminating against scattering from any slightly misaligned
secondary crystallites, and also provides for a well-defined
resolution in two dimensions in the scattering plane making
possible the mapping out of the intensity distribution of the
scattered x-rays in an area of reciprocal space. The set of
monochromator and analyser crystals is selected so as to meet
the resolution needs of a particular experiment but the
trade-off in intensity paid for using higher resolution also has
an important bearing on this choice. There are three sets of
crystals from which to choose, the properties and shape of the
resulting resolution function in reciprocal space have been
described for each in detail by Lucas [105]:
(1) Low resolution, using pyrolitic graphite crystals which
have a mosaic spread of
0.4
; these provide a maximum
of intensity and are ideally suited to measuring weak and very
diffuse features. They have been utilised in the bulk of
measurements described in this thesis.
(2) Medium resolution, using the 111 reflection from germanium
crystals with a mosaic spread of 4.3x10
degrees, the intensity
is reduced by a factor of
100 compared to graphite
but provides a resolution which is more than sufficient for
measuring the intrinsic widths of high-T
crystals.
(3) High resolution, which uses the (111) reflection of perfect silicon
crystals, limits the available intensity still further and therefore
ideally requires a scattering volume greater than the largest of
high-T
single crystals which can be currently grown. They
have not consequently been used in this work.
The sample crystal was fixed to a suitable holder which was
then screwed onto the goniometer head. The goniometer
has two axes of tilt and two of translation to allow
alignment of the sample in the beam. A sample was aligned
on the diffractometer in Bragg reflection geometry with its
reciprocal
axes in the horizontal scattering
plane. Because the incident beam is reduced to a size less
than that of the sample's face, it allows the elimination of
secondary crystallites by the translation of the sample across
the beam.